Compact thermometer with Pt100 or 4 to 20mA output

Temperature sensor with Pt100 or 4 to 20mA output

Easytemp TMR31

from €65.-
Need help?

Info-Hotline
2108002320
(εσωτ. 24)

    • PC programmable transmitter with 4 to 20 mA output
      or Pt100 sensor

    • Fast response times without reduced tip

    • Selectable sensor length

    Specs at a glance

    Temperature range –50 to +200 °C (–58 to 392 °F)
    Accuracy <0.08 %, Pt100 class A
    Immersion length (diameter)

    40 to 600 mm (Ø 6 mm)

    (1.6 to 23.6" (Ø 0.24))

    Sensor response time <1.0 s (T50), <2.0 s (T90)
    Product / VersionPrice / pcs. in EUROrder code
    Output; Measuring RangeProcess Connection1 to 34 to 1011+
    4..20mA max.150°W/o; compr. fitting, 316L, L=>100mm (3.94") insertion length Compr. fitting, to order separately.121,-112,-106,- TMR31-A1XxAAAX1AAA
    4..20mA max 150°Thread ISO228 G1/2, 316L128,-119,-113,- TMR31-A1XxBBAX1AAA
    4..20mA max.200°W/o; compr. fitting, 316L, L=>100mm (3.94") insertion length Compr. fitting, to order separately.121,-112,-106,- TMR31-A1XBAAAX1AAA
    4..20mA max 200°Thread ISO228 G1/2, 316L128,-119,-113,- TMR31-A1XBBBAX1AAA
    Pt100, DIN class A, 4-wireW/o; compr. fitting, 316L, L=>100mm (3.94") insertion length Compr. fitting, to order separately.74,-69,-65,- TMR31-A11xAAAX1AAA
    Thread ISO228 G1/2, 316L81,-75,-71,- TMR31-A11xBBAX1AAA

    Accessories

    Product / VersionPrice / pcs. in EUROrder code
    Collar welding boss d6 PEEK + screw moveable with sealing taper, washer + pressure screw G1/2, material of wetted parts: 316L, PEEK, max. Prozessdruck: 10 bar (145 psi)44,97 51004751
    Collar welding boss d6 PEEK w/o screw moveable with sealing taper + washer, material of wetted parts: 316L, PEEK, max. process pressure: 10 bar (145 psi)23,75 51004752
    Compression Fitting TA50
    6mm; G1/2; PTFE
    54,79 TA50-HP
    Configuration kit TXU10- for PC-programmable devices. set-up programme+interface cable for PC with USB-Port. 4 pin plug+ReadWin2000+adapter M12121,87 TXU10-BA
    Cable M12x1 l=5m 5m cable 4x 0.34 mm2, molded PVC angled connector, M12x1 connector screw Cu Sn/Ni, IP67.9,28 51005148

    Prices valid until 30.11.2017 in EUR per unit (not including shipping and applicable sales tax). See
    Endress+Hauser Terms and Conditions

    People who bought this also watched

  • Application

    The Easytemp TMR31 is a compact cost-effective thermometer that has been designed for use in all general applications across the process industries.

    Function

    The compact thermometer consists of a Pt100 class A sensor, (optional) transducer and housing with process connection. The TMR31 is available without a process connection so that it can be combined with a TW251 thermowell (with compression fitting) for insertion into a process. The built-in electronics can be programmed using a PC and TXU10-BA configuration kit to convert the Pt100 input signal into a temperature linear 4 to 20 mA signal. 

    Installation instructions

    Compact thermometer Easytemp TMR31 - INstalllation examples

    Pipe installation

     

    a) At angle sections, against the direction of flow

    b) In smaller pipes, turned against the direction of flow

    c) Perpendicular to the direction of flow

     

    L = Insertion length

  • Technical data

    Sensor
    Sensing element 1 × Pt100
    Tolerance Class A to IEC 751
    Process temperature

    –50 to +150 °C (–58 to 302 °F)

    (200 °C (392 °F) extension neck)

    Material 316L
    Diameter 6 mm (⌀ 0.24)
    Length 40 to 600 mm (1.6 to 23.6") selectable (please specify)
    Response time T50 = <1.0 s; T90 = <2.0 s
    Max. pressure 100 bar / 1500 psi (at 100 mm / 3.94" and 20 °C / 68 °F, depending on process conn.)
    Process connection
    Version G½", 316L; without (flange welding boss or compression fitting needed)
    Transmitter housing
    Degree of protection IP 66/67
    Electrical connection M12×1-plug-in connector
    Material 304
    Output characteristics
    Output signal With electronics 4 to 20 mA, 20 to 4 mA without electronics Pt100
    Signal on alarm
    • Under measuring range: linear decrease to 3.8 mA
    • Over measuring range: linear increase to 20.5 mA
    • Sensor break; Sensor short-circuit: ≤3.6 mA or 21.0 mA
    Load Max. (Upower supply –10 V )/0.023 A
    Induced current requirement ≤3.5 mA
    Current limitation ≤23 mA
    Switch-on delay 2 s
    Power supply
    Supply voltage Ub = 10 to 35 V DC
    Residual ripple Permitted residual ripple Uss ≤ 3 V at Ub ≥ 13 V, fmax = 1 kHz
    Accuracy
    Transmitter response time 1 s
    Reference operating conditions 0 °C (32 °F) (ice bath)
    Measuring error 0.1 K (0.18 °F) or 0.08 %
    Influence of supply voltage ≤ ±0.01 %/V deviation from 24 V
    Influence of ambient temperature (temperature drift)

    Td = ±(15 ppm/K × (full scale value of measuring range + 200) + 50 ppm/K × set measuring range) × ∆θ

    ∆θ = deviation of ambient temperature from the reference operating condition

    Long term stability ≤ 0.1 K/year or 0.05 %/year
    Influence of load ≤ ±0.02 %/100 Ω
    Operating conditions
    Ambient temperature –40 to +85 °C (–40 to 185 °F)
    Storage temperature –40 to +85 °C (–40 to 185 °F)
    Climate class As per EN 60 654-1, Class C
    Condensation Permitted
    Shock resistance 4 g/2 to 150 Hz as per IEC 60068-2-6
    EMC According to IEC 61326 Series and NAMUR NE21

    Dimensions (in mm)

    TMR31 without process connection

    Compact thermometer Easytemp TMR31 - Dimensions 1

    L1 = selectable length from 40 to 600 mm (1.6" to 23.6")

    * Version with extension neck add 35 mm (1.38")

    With G½" process connection

    Compact thermometer Easytemp TMR31 - Dimensions 2

    L1 = selectable length from 40 to 600 mm (1.6" to 23.6")

    * Version with extension neck add 35 mm (1.38")

    Electrical connection

    4 to 20 mA

    Compact thermometer Easytemp TMR31 : Electrical connection 1

    1 × Pt100

    Compact thermometer Easytemp TMR31 - Electrical connection 2

    a) red

    b) white

    Measuring range

    Code Measuring range

    AA

    4…20 mA, 0…+100 °C

    BA

    4…20 mA, 0…+150 °C

    CA

    4...20 mA, –50...+100 °C

    DA

    4...20 mA, –50...+150 °C

    EB

    4...20 mA, –50...+200 °C with extension neck
    FB 4...20 mA, 0...+200 °C with extension neck

    XB

    4...20 mA, measuring range selectable – please specify in  your order!

  • More documents for TMR31