Liquipoint FTW23 - Point level switch for liquids
A3EHEDGIOLINK



Point level switch for liquids in the food and beverage industry

Liquipoint FTW23

from €111.-
Need help?

Info Hotline
+31 (35) 6 95 86 11

    • Function testing of switch outputs with test magnet
    • Separate configuration of two switching thresholds, e.g. medium detection and medium differentiation
    • 3-A and EHEDG certificates

    Specs at a glance

    Product Water-based medium
    Installation Vessels and pipes
    Process temperature range

    –20 to +100 °C (–4 to +212 °F)

    (For 1 hour: +135 °C (+275 °F))

    Process pressure range –1 to +16 bar (–14.5 to +232 psi)
    Product / VersionPrice / pcs. in EUROrder code
    Power Supply; OutputElectrical ConnectionProcess Connection1 to 34 to 1011+
    10-30VDC; 3-wire PNP / IO-Link; DC-PNPPlug M12, IP65/67 NEMA Type 4 Encl.Thread ISO228 G1, 316L, install > accessory weld-in adapter147,-132,-120,- FTW23-AAxMWSJ
    Thread G1/2, 316L install > process sleeve hygienic142,-128,-116,- FTW23-AAxMWVJ
    Thread ISO228 G3/4, 316L, install > accessory weld-in adapter135,-122,-111,- FTW23-AAxMW5J
    Thread M24, 316L install > accessory adapter142,-128,-116,- FTW23-AAxMX2J
    Plug M12, IP66/68/69 NEMA Type 4/6P Encl.Thread ISO228 G1, 316L, install > accessory weld-in adapter169,-152,-139,- FTW23-AAxNWSJ
    Thread G1/2, 316L install > process sleeve hygienic165,-148,-135,- FTW23-AAxNWVJ
    Thread ISO228 G3/4, 316L, install > accessory weld-in adapter158,-142,-130,- FTW23-AAxNW5J
    Thread M24, 316L install > accessory adapter165,-148,-135,- FTW23-AAxNX2J

    Accessories

    Product / VersionPrice / pcs. in EUROrder code
    Weld-in adapter G3/4 d=50, 316L Flush mount process connection. For vessel-mounting. Material: 316L. Seal: silicone O-ring. FDA, EHEDG, 3-A compliant. Usage: FTL20/FTL20H, FTL31/FTL33, FTW23, FTW33, FMI51, FTI51, 11371. Usage: TMR35.25,77 71258355
    Weld-in adapter G3/4, d=29, 316L Flush mount process connection. For pipe-mounting. Material: 316L. Seal: silicone O-ring. FDA, EHEDG compliant. Usage: FTL20/FTL20H, FTL31/FTL33, FTW23, FTW33.26,- 71258357
    Weld-in adapter G1, d=60, 316L Flush mount process connection. Material: 316L. Seal: silicone O-ring. FDA, EHEDG, 3-A compliant. Usage: FMI51/FMI52, FTI51/FTI52, FTL20/FTL20H, FTL31/FTL33, FTL50/FTL50H/FTL51/FTL51C/FTL51H, FTL80/FTL81, FTL260, FTW23, FTW33, PMP23, PMP51, PMP135, PTP33B/PTP35. Usage: TM401/TM411, TMR35.37,76 52001051
    Weld-in adapter G1, d=53, 316L Flush mount process connection. Material: 316L. Seal: silicone O-ring. FDA, EHEDG, 3-A compliant. Usage: FMI51/FMI52, FTI51/FTI52, FTL20H, FTL31/FTL33, FTL50/FTL50H/FTL51/FTL51C/FTL51H, FTL80/FTL81, FTL260, FTL330H/FTL330L, FTW23, FTW33, PMP135, PTP35.37,77 71258358
    Cable 4x0.34, box M12, 316L,LED,90o,L=5m M12 socket for plug + integrated LED. 5m cable PVC, 316L lock. Especially for hygiene applications. Usage: device + switch output.20,22 52018763
    Cable 4x0.34, box M12,Cu Sn/Ni,90o,L=5m M12 plug connector. 5m cable PVC, Cu Sn/Ni lock. Usage: FMB50/FMB51/FMB52/FMB53, FTL31/FTL33, FTW23, FTW33, PMC11, PMC21, PMC51, PMP11, PMP21/PMP23, PMP51/PMP55, PTC31B, PTP31B/PTP33B.7,05 52010285
    Socket for plug 4p-125V, 4A-M12 M12 socket for plug. Self confectionery connection. Usage: plug M12.15,32 52006263
    Test magnet7,05 71267011

    Prices valid until 12.31.2018 in EUR per unit (not including shipping and applicable sales tax). See
    Endress+Hauser Terms and Conditions

    Is this Liquipoint level switch not the perfect match for your application?

    People who bought this also watched

  • Application

    The Liquipoint FTW23 is a point level switch for water-based liquids. It is used preferably in storage tanks, mixing vessels and pipes. Developed and built for the food and beverage industry, the Liquipoint FTW23 meets international hygienic requirements. The Liquipoint FTW23 can be used permanently in process temperatures up to 100 °C (212 °F) and in cleaning and sterilization processes to 135 °C (275 °F) for 60 minutes.

    Function

    The capacitance at the tip of the sensor, and therefore the dielectric value of the medium, is determined using an electrical field. Given that the dielectric value of air and a water-based liquid differ, the Liquipoint FTW23 can differentiate between the two states, i.e. covered and uncovered.

    Application Example

    Liquipoint FTW23 - Point level switch - Application Example

    The measuring system consists of a Liquipoint FTW23 point level switch, e.g. for connection to programmable logic controllers (PLC).

    a) Overfill protection or upper level detection (MAX)

    b) Pump dry running protection (MIN)

    c) Lower level detection (MIN)

  • Technical data

    Output
    Switch output 3-wire DC-PNP:
    – 2 DC-PNP outputs, switched using XOR operation
    – 200 mA connectable load (short-circuit proof)
    Devices with IO-Link:
    – 2 DC-PNP outputs, freely configurable
    – 1 switch output active: 200 mA connectable load (short-circuit proof)
    – Both switch outputs active: connectable load of 105 mA each (short-circuit proof)
    Residual voltage <3 V
    Residual current <100 μA
    Supply voltage 10 to 30 V DC
    Power consumption <1.2 W (at max. load: 200 mA)
    Current consumption <40 mA
    Cable specification IEC 60947-5-2
    Connecting cable
    length
    – max. 25 Ω/core, total capacity < 100 nF
    – IO-Link communication: < 10 nF
    Performance characteristics
    Reference operating
    conditions

    Horizontal orientation:

    – Ambient temperature: 20 °C (68 °F) ±5 °C

    –Medium temperature: 20 °C (68 °F) ±5 °C

    – Process pressure: 1 bar (14.5 psi)

    –Medium: water

    Switching accuracy ±2 mm (0.08 in) in accordance with DIN 61298-2
    Hysteresis Typically ±1 mm (0.04 in)
    Non-repeatability ±1 mm (0.04 in) in accordance with DIN 61298-2
    Switching delay

    0.5 s when sensor is covered

    1.0 s when sensor is uncovered

    Switch-on delay <2 s (previously not through-connected)
    Orientation any position
    Environment
    Ambient temperature
    range

    –20 to +70 °C (–4 to +158 °F) (at TProcess ≤ 80 °C (176 °F)),

    –20 to +35 °C (–4 to +95 °F) (at TProcess = 135 °C (275 °F))

    Storage temperature –40 to +85 °C (–40 to +185 °F)
    Climate class DIN EN 60068-2-38/IEC 68-2-38: test Z/AD
    Degree of protection

    – IP65/67 NEMA Type 4X Enclosure (M12 connector for plastic housing cover)

    – IP66/68/69K NEMA Type 4X/6P Enclosure (M12 connector for metal housing cover)

    Short-circuit
    protection

    – Overload protection/short-circuit protection at I > 200 mA
    – Device with IO-Link: 105 mA per output if both switch outputs are active

    Process
    Process temperature
    range
    –20 to +100 °C (–4 to +212 °F) (For 1 hour: +135 °C (+275 °F))
    Process pressure
    range
    –1 to +16 bar (–14.5 to +232 psi)
    Process fluid – Water-based media with a dielectric constant (DC) > 20 (default)
    – Device with IO-Link communication: adjustment up to DC > 1.5 via the IO-Link interface for water-, alcohol- and oil-based liquids or powdered products
    Mechanical construction
    Weight Max. 300 g (10.58 oz)
    Materials in contact
    with process

    – Sensor: 316L (1.4404), PEEK The material PEEK meets the requirements of EU 1935/2004, 10/2011,
    2023/2006 and FDA 21 CFR 177.2415

    – Process connection: 316L (1.4404/1.4435)

    Materials not in
    contact with process

    Housing covers:

    – M12 metal: 316L (1.4404)

    – M12 plastic: PPSU; Design ring: PBT/PC; Housing: 316L (1.4404)

    Surface Ra ≤ 0.76 μm (30 μin)
    Approvals
    Approval CSA C/US General Purpose
    Sanitary compatibility 3-A, EHEDG

    Dimensions in mm (inches)

    Housing, electrical connection

    Liquipoint FTW23 - Point level switch – Dimensions

    a) M12 (metal)

    b) M12 (plastic)

     

    Installation according to instruction manual

     

    Process connections

    Liquipoint FTW23 - Point level switch – Process connections

    a) G¾"

    b) G1"

    c) M24×1,5

    d) G½" Hygiene adapter

     

    Installation according to instruction manual

     

    Electrical connection

     

    Liquipoint FTW23 - Point level switch – Electrical connections

    a) M12 connector

    b) Function monitoring with antivalence

    c) IO-Link with one switch output

  • More documents for FTW23
  • IO-Link

    Besides its general benefits, as summarized below, using IO-Link point-to-point communication with our hygienic level switch Liquipoint FTW23 allows:


    • Setting up a second switch output

    • Setting up the device for usage in products with DC values down to 1,5 (non-conductive media)

    • Setting up the device for usage in powdery products (e.g. coffee powder)

     

    General benefits of IO-Link

    Digital measured values:

    • Faultless transmission (e.g. EMC caused)

    • No conversion, thus a higher total accuracy

     

    More data

    • Possibility to carry more values

    • Detailed fault information

    • Additional process information

     

    Central parameters storage

    • Possibility to load set up parameters from the master into a device

    • Configuration can be duplicated from one device to the next

    • Remote sensor maintenance